Product Details
Product Specifications
Assay TypeSandwich-ELISAAnalyteAAV9Format96TReactivity/Regulatory StatusRUOSensitivity≤100 ng/mLStandard Curve Range0.313 ng/mL-20 ng/mLAssay Time3 hr 20 minSuitable Sample TypeFor the quantitative determination of anti-AAV9 antibodies in serum.Sample volume100 μLProduct Overview
Anti-AAV9 Antibody ELISA Kit is based on the sandwich-ELISA method and is used to detect and quantitatively determine the Anti-AAV9 Antibody in serum. The kit is designed to provide reliable solutions for the immunogenicity studies and enrollment screening of AAV gene therapy products.
Reconstitution
Please see Certificate of Analysis for details of reconstitution instruction and specific concentration.
Storage
1. Store the unopened kit at 2-8°C upon receipt.
2. Locate the expiration date on the outer packaging and do not use reagents beyond their expiration date.
3. The opened kit should be stored according to the storage conditions listed in the Components Table. The shelf life of the opened kit is 30 days from the date of opening.
Materials Provided
IDComponentsSizePAV009-C01Pre-coated AAV9 Capsid Protein Microplate1 platePAV009-C02Anti-AAV9 Antibody Standard2.5 μgPAV009-C03Biotin-AAV9 Capsid Protein10 μgPAV009-C04Streptavidin-HRP10 μgPAV009-C0510xWashing Buffer50 mLPAV009-C062xDilution Buffer50 mLPAV009-C07Substrate Solution12 mLPAV009-C08Stop Solution7 mLACRO Quality Management System
Customers Also Viewed
Performance Data
Typical Data
Please refer to DS document for the assay protocol.

Detection of anti-AAV9 antibodies by sandwich-ELISA Assay.
Immobilized AAV9 Capsid Protein can bind Anti-AAV9 Antibody in 1:50 human serum, and then add Biotin-AAV9 Capsid Protein. Detection was performed using Streptavidin-HRP with sensitivity of 15.6 ng/mL.
The specific binding of PAV-A009 to Anti-AAV3/5/8 antibody was detected. The results showed that PAV-A009 only binds specifically to Anti-AAV9 antibody.
Intra-Assay Statistics
Three samples of known concentration in blank human serum were tested six times on one plate to assess intra-assay precision, Intra-Assay Precision CV≤15%.

Three samples of known concentration in blank mouse serum were tested six times on one plate to assess intra-assay precision, Intra-Assay Precision CV≤15%.

Inter-Assay Statistics
Three samples of known concentration in blank human serum were tested in six separate assays to assess inter-assay precision, Inter-Assay Precision CV≤15%.

Three samples of known concentration in blank mouse serum were tested in six separate assays to assess inter-assay precision, Inter-Assay Precision CV≤15%.

Customer Reviews Writing Reviews